[. . . ] Installation and Use of 32mm molbloc Cradle Kit P/N 402134 Instruction Sheet This kit is used to retrofit the 1/4" molstic-STM platform in order to provide mounting for the molbloc-LTM laminar flow elements with 32 mm square bodies. PREPARING THE PLATFORM FOR CRADLE INSTALLATION Early deliveries of the molstic-S platform did not include the holes necessary to mount the downstream cradles for 32 mm molbloc-L elements. [. . . ] For example, if installing the downstream cradle in location "4", mark a line from the centers of the two holes in location "7" to the centers of those in location "1". Punch the intersection of the marked lines to facilitate drilling the platform. Drill all necessary holes to a diameter of 4. 6 mm. INSTALLATION OF CRADLES ONTO MOLSTIC-S PLATFORM With the molstic-S resting on its side, position the upstream cradle for the 32 mm molbloc-L into the desired location, either 1, 2, or 3. Place a lock washer and then a flat washer onto the socket head cap screw. Thread the screw into the upstream cradle from the bottom side of the molstic platform and leave loose enough to allow for later adjustment. Thread another screw with lock washer and flat washer into the other side of the upstream cradle. Position the downstream cradle for the 32 mm molbloc-L into the desired location, either 4, 5, or 6. Repeat these step with any other cradles that might need to be installed. [. . . ] Document 560070 040907 DH Instruments, Inc. 4765 East Beautiful Lane Phoenix AZ 85044-5318 USA Tel 602. 431. 9100 Fax 602. 431. 9559 dhi@dhinstruments. com www. dhinstruments. com [. . . ]