[. . . ] Instruction Manual
HI 9023C - HI 9023CN HI 9024C - HI 9025C HI 9110 - HI 9210 HI 9210N Portable Waterproof pH Meters
Dear Customer, Thank you for choosing a Hanna Product. Please read this instruction manual carefully before using the instrument. It will provide you with the necessary information for a correct use of the instrument, as well as a precise idea of its versatility. If you require further technical information, do not hesitate to e-mail us at tech@hannainst. com. [. . . ] With the temperature probe located
CAL
1
CAL
BUF
pH
as close to it as possible.
CAL
TWO-POINT CALIBRATION After calibrating at 6. 86 or 7. 01 pH (see above), for the second buffer use pH 4. 01 if you are going to measure acidic samples, and pH 9. 18 or pH 10. 01 if you are measuring alkaline solutions. · Proceed as described in "One-point calibration above but do not end the calibration by pressing CAL. · After the first calibration point is confirmed, immerse the pH electrode into the second buffer solution (pH 4. 01, 9. 18 or 10. 01) and stir gently. Note: The electrode should be submerged approximately 4 cm (1½") into the solution with the temperature probe located as close to it as possible. · Select the second buffer value on the or secondary display by pressing the °C or C key.
HI 7004
°C
1
BUF
°C
4 cm (1½")
°C
pH
19
20
°C
Note: The electrode should be submerged approximately 4 cm (1½") into the solution with the temperature probe located as close to it as possible. · Select the second buffer value on the secondary display by
READY CON
· Press pH/ C SET key for HI 9110 or the RANGE key for HI 9210 to select the manual temperature setting. The " C" symbol will be displayed to indicate the temperature setting mode.
WRONG WRONG
CFM
BUF
2
°C
pressing the °C or C key. · When the "READY" and "CON" symbols blink on the display, the reading is stable and the calibration can be confirmed. If the reading is not close to the selected buffer solution, "WRONG 2 " and "WRONG " will blink alternatively. If the reading is close to the selected buffer, the slope and the offset are calibrated. The values will be stored in memory and the meter will return to the operating mode. Note: The meter will automatically skip the buffer that was used for the first calibration to avoid erroneous calibration. At least 1 full pH unit is required between the two buffers used for the offset and slope calibration. As a result, once calibrated at either pH 6. 86 or 7. 01, the microprocessor will automatically ignore the other one for the second point calibration. It will do likewise for pH 9. 18 and 10. 01 Note: During calibration, the secondary LCD displays the selected buffer value. With HI 9025C, press the RANGE key to display the buffer temperature during calibration.
BUF
· Remove the protective cap on the top of the meter to have access to the calibration trimmers.
pH °C SET
RANGE
PROCEDURE FOR HI 9110, HI 9210 & HI 9210N:
Manual Temperature Compensation Note: The pH/ C electrode of HI 9210N incorporates a temperature sensor which performs automatic temperature
pH °C SET
°C
· Record the temperature of the buffer with a Checktemp C or an accurate thermometer (e. g. 20 C). · Using the screwdriver, turn the temperature trimmer (# 5 pages 9-10) to display the recorded temperature value. · Press the pH/ C SET key for HI 9110 or the RANGE key for HI 9210 to read pH. As a result, you can perform one or two point calibration at a buffer of your choice. The instructions below explain a typical calibration procedure at pH 7 and then 4 or 10. · Remove the protective cap from the electrode, rinse it with some pH 7. 01 solution, then dip in pH 7. 01 buffer. Note: The electrode should be submerged approximately 4 cm
°C
HI 7007
RANGE
compensation. [. . . ] Use: general purpose.
12 mm 0. 5"
HI 1413B Glass-body, single junction, flat tip, Viscolene, non-refillable, combination pH electrode. Use: surface measurement.
12 mm 0. 5"
HI 4430 120 mm 4. 7"
"S" VERSION
Consult the Hanna General Catalog for an extensive selection of electrodes.
HI 1413
110 mm 4. 3"
OTHER ACCESSORIES
ChecktempC Pocket-size thermometer (range -50. 0 to 150. 0 C) HI 76405 Electrode holder HI 7669/2W Temperature probe with 1 m (3. 3') screened cable (not HI 9110, HI 9210, HI 9210N and HI 9023CN) HI 8427 pH and ORP electrode simulator with 1 m (3. 3') coaxial cable ending in female BNC connectors HI 931001 pH and ORP electrode simulator with LCD and 1 m (3. 3') coaxial cable ending in female BNC connectors HI 721317 Rugged carrying case 36
Ultem® is registered Trademark of "General Electrics Co. " Kynar® is registered Trademark of "Pennwalt Corp. "
35
ELECTRODE APPLICATION REFERENCE GUIDE
Application 1. Aquarium Bath-water Beer Bread Cheese Dairy products Dirty water Emulsions Environment Flasks Food industry general use Fruit Fruit juices, organic Galvanizing waste solution Heavy-duty applications High purity water Horticulture Laboratory general use Leather Lemon juice Meat Micro plate sampling of less than 100 mL Milk and Yogurt Paints Paper Photographic chemicals Quality control Sausages Semisolid products Skin Soil samples Solvents Strong acid Submersion application Surface measurements Swimming pool Titrations with constant temperature range Titrations with wide temperature range Very high humidity Vials and test tube Wine processing Electrodes * HI 1332B HI 1130B HI 1131B HI 2031B, FC 200B FC 200B FC 100B HI 1230B HI 1053B HI 1230B HI 1331B FC 911B, FC 100B FC 200B, FC 220B FC 210B HI 1130B HI 1135B HI 1053B HI 1053B, FC 200B HI 1131B, HI 1230B, HI 1332B, HI 1330B HI 1413B FC 100B FC 200B, HI 2031B HI1083B FC 210B HI 1053B HI 1413B HI 1230B HI 1332B FC 200B, HI 2031B HI 2031B HI 1413B HI 1230B HI 1043B HI 1043B HI 1130B HI 1413B H I1130B, HI 2114P/2 HI 1131B HI 1131B FC 911B HI 1330B FC 220B
WARRANTY
All Hanna Instruments meters are guaranteed for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. The electrodes and the probes are guaranteed for a period of six months. [. . . ]