[. . . ] Checktemp2
Pocket-sized Thermometer
FRONT
ON/OFF SWITCH
BACK
SIDE
BATTERY CASE
LIFT THE POCKET CLIP HERE TO RELEASE THE PROBE LIQUID CRYSTAL DISPLAY POCKET CLIP
180° MAXIMUM
IMMERSION LEVEL TEMPERATURE PROBE
OPERATION: · Turn the Checktemp 2 on
by pressing the ONOFF switch located on the front. [. . . ] When not in use, switch Checktemp 2 off.
°c
BATTERY REPLACEMENT: When the Checktemp 2 cannot be
switched on or the display fades, pull out the battery compartment (after unscrewing the screw underneath the battery clip) and change the 1. 4V battery, paying attention to the polarity.
BATTERY COVER
·
·
°c
°C
BATTERY
·
LOOSEN SCREW POCKET CLIP (LIFTED)
·
Batteries should only be replaced in a safe area using the battery type specified in this instruction manual.
CALIBRATION CHECK DEVICE (CCD):
Beneath the battery compartment cover there is a CCD switch. When activated, the circuit will send a signal to the LCD requesting a 0. 0°C (32. 0°F) reading. The Checktemp 2 should display 0. 0°C ±0. 3°C (32°F ±0. 5°F). This assures the user that the reading is reliable and accurate.
CALIBRATION CHECK MODE MEASUREMENT MODE
RECOMMENDATIONS FOR USERS:
Before using this product, make sure that it is entirely suitable for the environment in which it is used. Operation of this instrument in residential areas could cause interference to radio and TV equipment. Any variation introduced by the user to the supplied equipment may degrade the instrument's EMC performance. [. . . ] To avoid damage or burns, do not perform any measurement in microwave ovens.
BATTERY COVER
CALIBRATION CHECK DEVICE LOOSEN SCREW POCKET CLIP (LIFTED)
Checktemp2
Pocket-sized Thermometer
SPECIFICATIONS:
Checktemp 2 C Checktemp 2 F
RESOLUTION RANGE -50. 0 to 150. 0°C -58. 0 to 302°F 0. 1°C
Checktemp 2 C Checktemp 2 F 0. 1°F Checktemp 2 C Checktemp 2 F Checktemp 2 C Checktemp 2 F
ENVIRONMENT
(-58. 0 to 199. 9°F) 1°F (200 to 302°F)
ACCURACY (@20°C/68°F)
±0. 3°C (-20 to 90°C) ±0. 5°C (outside) ±0. 5°F (-4 to 194°F) ±1°F (outside)
TYPICAL EMC DEVIATION ±0. 3°C ±0. 5°F 0 to 50ºC (32 to 122ºF) 95% RH
BATTERY TYPE 1 x 1. 4V alkaline LIFE approx. 3000 hours PROBE DIMENSIONS WEIGHT Folding, 110mmx3mm diameter 150x30x20 mm (5. 9x1. 2x0. 8") 60 g (2. 1 oz. )
ACCESSORIES:
HI76504/P1 1. 4V battery
Visit our Internet Home Page: http://www. hannainst. com
Authorized Dealer
ISTRCHT2 01/97 [. . . ]