[. . . ] Before beginning all unknown measurements, set meter to highest possible range. Before breaking a circuit for testing, turn off the power to the circuit. When disconnecting from a circuit, disconnect the hot lead first, then the common lead. Disconnect the meter from the circuit before turning off any indicator, including motors, transformers, and solenoids.
2
2 1 5
3 4
1. [. . . ] To disable the auto power off function, power up the meter while pressing either the range, RS-232 or min/max. button.
4
Overload Protection Function VAC & VDC AAC & ADC Ohms () Diode Continuity Overload Protection 1000V A input: 1A(600V) fast blow fuse µA input: 600V RMS 600VAC/600VDC 600VAC/600VDC 600VAC/600VDC
Unit of Measure Multipliers For your reference, the following symbols are often used to make measurement easier: Symbol Verbal Multiplier M mega x 1, 000, 000 k kilo x 1, 000 m milli ÷ 1, 000 µ micro ÷ 1, 000, 000
5
True RMS AC Volt Circuit Connection:
Range 600mV 6V 60V 600V 1000V
Resolution 100µV 1mV 10mV 100mV 1V
Max. Display 600. 0 ±(0. 9% reading 6. 000 + 5 digits) 60. 00 600. 0 1000
Accuracy
AC Conversion Type: AC conversions are AC-coupled, true RMS responding, calibrated to the RMS value sine wave input. Plug the black test lead into the COM port and the red test lead into the Hz V µA port. Measure AC Voltage.
6
DC Volts Circuit Connection:
Range 600mV 6V 60V 600V 1000V
Resolution 100µV 1mV 10mV 100mV 1V
Max. Display 600. 0 ±(0. 5% reading 6. 000 + 2 digits) 60. 00 600. 0 1000
Accuracy
Input Impedance: 10M (over 1000M in 600mV range) To Measure DC Voltage: 1. Plug the black test lead into the COM port and the red test lead into the Hz V µA port. Measure DC Voltage
7
True RMS AC Current Circuit Connection:
Range Resolution 60mA 600mA 10µA 100µA
Accuracy +(1. 5% reading +5
Voltage Burden 2V max
Max 60. 00 600. 0
AC Conversion Type: AC conversions are AC-coupled, true RMS responding, calibrated to the RMS value sine wave input. Overload Protection: A input: 1A (600V) fast blow fuse Crest Factor: +1. 5% additional error for C. F. from 3 to 4 Frequency Response: 40 Hz ~ 500Hz To Measure True RMS AC Current: 1. Plug the black test lead into the COM port and the red test lead into the mA. 8
DC Current Circuit Connection:
Range Resolution 600µA 6000µA 60mA 600mA 0. 1µA 1µA 10µA 100µA
Voltage Max Burden < 4mV / µA 600. 0 ±(1. 0% reading 6. 000 + 2 digits) 2V max 60. 00 600. 0
Accuracy
Overload Protection: A Input: (600V) fast blow fuse µA Input: 600V RMS To Measure DC Current: 1. Plug the black test lead into the COM port and the red test lead into the mA or Hz V µA port. Measure DC Current.
9
Frequency/RPM Circuit Connection:
Range
6000Hz 60KHz 600KHz 6MHz 60MHz
Resolution
1Hz 10Hz 100Hz 1KHz 10KHz
Sensitivity
100mV RMS
Accuracy
Frequency: 0. 01% ±1 digit
250mV RMS 1V RMS
Overload Protection: 600V rms *Less than 20Hz the sensitivity is 1. 5V To Measure Frequency: 1. Plug the black test lead into the COM port and the red test lead into the Hz V µA port. Measure Frequency or RPM.
10
Resistance (Ohms) Circuit Connection:
Range 600 6K 60K 600K 6M 60M
Resolution Accuracy Max. Display 0. 1 600. 0 1 ±(0. 7% reading 6. 000 10 + 2 digits) 60. 00 100 600. 0 1 6. 000 10 60. 00
Open Circuit Voltage: -1. 3V approx. Turn the power off to the circuit or device that is to be measured and discharge all capacitors before attempting a measurement. Plug the black test lead into the COM port and the red test lead into the Hz V µA port. For correct reading, ensure that the device being tested contains no voltage. 11
Multiplication Guide for Ohms () 400 = Meter indicates actual resistance 4k = Multiply meter display reading by 1, 000 to acquire actual resistance. 40k = Multiply meter display reading by 1, 000 to acquire actual resistance. [. . . ] Plug the black test lead of the clamp adapter handle into the COM port and the red test lead into the Hz V µA port. Snap the jaw of the current clamp around one of the current carrying conductors. Additional Current Clamps: 200AAC 61-332 500 AAC 61-333 600 AAC/DC 61-334 1000 AAC 61-436 Flexible Clamp Adapters: 1000 AAC 61-330 3000 AAC 61-31
17
General Specifications LCD Display: Polarity Indication: 6000 count maximum reading Automatic, negative indicated, positive implied Overrange Indication: "OL" or "-OL" Low Battery Indication: " " when the battery voltage drops below operating voltage Size (WxHxD): 82mm x 164mm x 44mm (without holster) Sampling: 1 time/sec LCD Display, Auto Power Off: Approx. Operating Temperature: 0°C ~ 30°C (<80% RH), 30°C ~ 40°C (<75% RH), 40°C ~ 50°C (<45% RH) Storage Temperature: -20°C ~ 60°C (0~80% RH) when battery removed from meter Temperature Coefficient: 0. 15 x (specified accuracy) / °C, <18°C or >28°C Power Requirements: 9V NEDA 1604, 1EC bf 22, J1S 006P Battery Life: 300 hours (alkaline ) Installation Category: IEC 1010, 1000V Cat. [. . . ]