Detailed instructions for use are in the User's Guide.
[. . . ] Alpha pH 800
Controller/Transmitter pH / ORP
ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perpHect, PerpHecT, PerpHecTion, pHISA, pHuture, Pure Water, Sage, Sensing the Future, SensorLink, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS and TURBO2 are registered trademarks of Thermo Fisher. 1-888-pHAX-ION, A+, All in One, Aplus, AQUAsnap, AssuredAccuracy, AUTO-BAR, AUTO-CAL, AUTO DISPENSER, Auto-ID, AUTO-LOG, AUTO-READ, AUTO-STIR, AutoTest, BOD AutoEZ, Cable-Free, CERTI-CAL, CISA, DataCOLLECT, DataPLUS, digital LogR, DirectCal, DuraProbe, Environmental Product Authority, Extra Easy/Extra Value, FAST QC, GAP, GLPcal, GLPcheck, GLPdoc, ISEasy, KAP, LabConnect, LogR, Low Maintenance Triode, Minimum Stir Requirement, MSR, NISS, One-Touch, One-Touch Calibration, One-Touch Measurement, Optimum Results, Orion Star, Pentrode, pHuture MMS, pHuture Pentrode, pHuture Quatrode, pHuture Triode, Quatrode, QuiKcheK, rf link, ROSS Resolution, SAOB, SMART AVERAGING, Smart CheK, SMART STABILITY, Stacked, Star Navigator 21, Stat Face, The Enhanced Lab, ThermaSense, Triode, TRIUMpH, Unbreakable pH, Universal Access are trademarks of Thermo Fisher. Guaranteed Success and The Technical Edge are service marks of Thermo Fisher. [. . . ] After leaving Calibration mode, controller switches back to process temperature (for setting calibration temperature and process temperature, see section 7. 2. 3).
14
6. 2 ORP mV Calibration This mode allows one-point calibration. Note: If upper and lower display read "CAL PH", see section 7. 4 for procedures on how to switch from pH to ORP mV readings. Upper display shows current mV output of electrode without any offset adjustment. The controller calculates the correction factor and returns to Measurement mode.
HOLD CAL SETUP
HOLD
CAL
OrP
226
C
mV
15
7
7. 1
ADVANCED SET UP MODE
Electrode Offset (OFS) sub-function
This option is available only in the pH mode. An allowance of ±120mV is allowed to correct for electrode deviations. This feature is useful for prolonged on-line applications.
The controller will add or subtract the value from the measured pH and display the correct value. Should the offset value be too high, consider performing maintenance on the electrode followed by calibration. If the calibration is unsuccessful, please replace the electrode.
SETUP
1) Take a sample from the system and note pH reading
HOLD
Ofs
displayed on controller. 2) Measure pH of sample using a calibrated pH tester, hand-held meter, or bench meter. Record the correct pH
SETUP
value.
HOLD
-0. 05
7. 00
3) Enter Advanced set-up mode. 4) Press or keys to scroll until upper display shows OFS.
5) Press ENTER key. 6) Press or keys until pH value coincides with correct pH value noted in Step 2. As the or keys is pressed to adjust pH value, the `offset' value in upper display also changes. 7) Press ENTER key to enter offset value.
16
Note: Offset value is reset during full calibration. Note: The offset parameter is blanked out during ORP operation.
7. 2
7. 2. 1
Setting temperature (Set oC) sub-function
Selecting automatic or manual temperature compensation
1) Enter Advanced Set-up mode by pressing ENTER key. The lower display shows "Atc"; upper display shows "on" or "oFF" depending on whether or not ATC is selected. 5) Proceed with additional Advanced Set-up
SETUP SETUP HOLD SETUP HOLD
SEt
o
C
Atc
on
procedures (press ENTER again) or return to Measurement mode by pressing and keys (escape) simultaneously.
HOLD
oFF
Atc
17
7. 2. 2
Temperature calibration (ATC mode only)
1) Select "ATC on" as described above in Section 7. 2. 1. The
SETUP HOLD
on
Atc
current measured temperature is shown in the lower display. 3) Compare current measured temperature on the controller display to a thermometer known to be accurate. 4) Press or keys to scroll the lower display to match the correct value. 6) Continue with additional Advanced Set-up procedures, or return to the Measurement mode by pressing and keys (escape) simultaneously.
SETUP HOLD
0. 0
25. 0 ATCC
o
18
7. 2. 3
Setting manual temperature compensation
SETUP
Note: This option is not available when the
HOLD
oFF
Atc
controller is set for ATC operation.
For manual temperature compensation, two
SETUP
different temperatures: process and calibration,
HOLD
25. 0
can
be
input
independently.
This
allows
P. C
calibration at a temperature other than the process temperature. Example:
o
setting
a
calibration temperature of 25 C allows calibration
SETUP HOLD
25. 0
C. C
using standard buffer solutions at 25oC, even if the process temperature is different.
1) Select "ATC oFF" as described above in section 7. 2. 1 2) Press ENTER key. [. . . ] Ryton-body pH Combi Electrode with Pt 100 RTD (ATC) & 5 m cable with BNC & connector for PMP & Pt 100. Ryton-body pH Combi Electrode without ATC & 5 m cable with BNC connector. Ryton-body ORP Platinum Electrode with 5 m cable with BNC & connector for PMP (No ATC). Ryton-body ORP Gold Electrode with 5 m cable with BNC & connector for PMP (No ATC). [. . . ]